The HC2S3, manufactured by Rotronic, is a rugged, accurate temperature and humidity probe that is ideal for long-term, unattended applications. The probe uses an advanced capacitive sensor to measure relative humidity. The probe includes a filter to protect the sensor from dust and particles, for superior performance and reliability.
Read More
The HC2S3 comes with a polyethylene filter that protects its sensor from fine dust and particles and minimizes water absorption and retention. Alternatively, a Teflon filter is available for marine environments. The response time is slower when using the Teflon filter.
Please note: The following shows notable compatibility information. It is not a comprehensive list of all compatible products.
| Product | Compatible | Note |
|---|---|---|
| 21X (retired) | ||
| CR10 (retired) | ||
| CR1000 (retired) | ||
| CR1000X (retired) | ||
| CR10X (retired) | ||
| CR200X (retired) | ||
| CR206X (retired) | ||
| CR211X (retired) | ||
| CR216X (retired) | ||
| CR23X (retired) | ||
| CR295X (retired) | ||
| CR3000 (retired) | ||
| CR500 (retired) | ||
| CR5000 (retired) | ||
| CR510 (retired) | ||
| CR800 (retired) | ||
| CR850 (retired) | ||
| CR9000 (retired) | ||
| CR9000X (retired) |
The HC2S3 should be housed in a 41003-5 10-plate naturally-aspirated solar radiation shield. This radiation shield mounts to a tripod or tower mast, crossarm, or user-supplied pole.
| Electronics Operating Limits | -40° to +100°C |
| Storage Temperature Range | -50° to +100°C |
| Current Consumption |
|
| Supply Voltage | 5 to 24 Vdc |
| Startup Time |
1.5 s (typical) The startup time is a Rotronics specification. For programming, Campbell Scientific recommends a 2 s delay at 60°C, a 3 s delay at 0°C, and a 4 s delay at -40°C. |
| Maximum Startup Current | < 50 mA (for 2 μs) |
| Analog Outputs |
|
| Diameter | 15 mm (0.6 in.) |
| Length |
|
| Weight | 10 g (0.35 oz) |
Air Temperature |
|
| Measurement Range |
-40° to +60°C (default) Air temperature measurement ranges from -50° to +100°C are possible; contact Campbell Scientific for more information. |
| Output Signal Range | 0 to 1 V |
| Long-Term Stability | < 0.1°C/year |
| Sensor Time Constant - Standard PE Filter | ≤ 22 s (63% step change [1 m/s air flow at sensor]) |
| Sensor Time Constant - Optional Teflon Filter | ≤ 30 s (Typical 4 s, 63% of a step change [1 m/s air flow at sensor]) |
| Temperature Accuracy | Refer to graph in probe manual. |
Relative Humidity |
|
| Measurement Range | 0 to 100% RH (non-condensing) |
| Output Signal Range | 0 to 1 Vdc |
| Long-Term Stability | < 1% RH per year |
| Accuracy | ±0.8% RH with standard configuration settings (at 23°C) |
| Sensor Time Constant - Standard PE Filter | ≤ 22 s (63% of a 35 to 80% RH step change [1 m/s air flow at sensor]) |
| Sensor Time Constant - Optional Teflon Filter | ≤ 30 s (Typical 10 s, 63% of a 35 to 80% RH step change [1 m/s air flow at sensor]) |
| RH Accuracy over Temperature | Refer to graph in probe manual. |