The CSIM11-ORP, manufactured by Wedgewood Analytical, measures oxidation reduction potential (ORP) of liquids. It can be submerged in water or inserted into tanks, pipelines, and open channels. The CSIM11-ORP is intended for non-pressurized systems and was not designed for applications above 30 psi. Please contact Campbell Scientific for recommendations on probes suitable for installations in pressurized pipes.
Read MoreThe CSIM11-ORP is similar to the CSIM11, but includes a 0.2-in. platinum band wrapped around the glass electrode, which allows the CSIM11-ORP to respond to the electron density in the fluid. It has a plunger-style pH glass electrode that allows the CSIM11-ORP to be mounted at any angle. The probe's porous polytetrafluoroethylene (PTFE) liquid junction is less susceptible to clogging as compared to conventional reference junctions.
A titanium ground rod runs inside the PPS outer body to eliminate ground loop errors. An internal amplifier boosts the signal, decreasing signal interference. The amplifier is powered by two internal lithium batteries, and thus does not require any current from the datalogger. These batteries are designed to last the lifetime of the probe.
The reference solutions and bulb configuration are optimized for natural water applications. Alternate reference solutions and bulb configurations are available. Contact Campbell Scientific for pricing and availability.
Note: The CSIM11-ORP uses glass bulb technology which has a life expectancy of around 6 months to 2 years, depending on the conditions of the water.
Output | Analog |
ORP Range | -700 to +1100 mV |
Operating Temperature Range | 0° to 80°C |
Pressure Range | 0 to 30 psig |
Accuracy | ±0.1% (over full range) |
Impedance | < 1 Mohm (@ 25°C) |
Reference Cell | Single Junction KCl/AgCl |
Body Material | ABS |
Wetted Materials | ABS, polytetrafluoroethylene (PTFE), Viton, glass, titanium |
Cable Jacket Material | Polyurethane |
Response Time | 95% of reading (in 10 s) |
Drift | < 2 mV per week |
Internal Lithium Battery Lifetime | 5 years (life of probe) |
Diameter | 3.0 cm (1.2 in.) |
Length | 17.8 cm (7.0 in.) |
Weight | 0.5 kg (1 lb) with 4.57-m (15-ft) cable |
Please note: The following shows notable compatibility information. It is not a comprehensive list of all compatible products.
Product | Compatible | Note |
---|---|---|
CR1000 (retired) | ||
CR1000X | ||
CR300 | ||
CR3000 (retired) | ||
CR310 | ||
CR350 | ||
CR6 | ||
CR800 (retired) | ||
CR850 (retired) |
The CSIM11-ORP requires one differential analog input channel.