HFP01SC-L Self-Calibrating Soil Heat Flux Plate
Extreme Accuracy
Self-calibrating; ideal for energy-balance in eddy-covariance and Bowen-ratio systems
weather aplicaciones soportadas water aplicaciones soportadas energy aplicaciones soportadas gas flux and turbulence aplicaciones soportadas infrastructure aplicaciones soportadas soil aplicaciones soportadas

Overview

The HFP01SC, manufactured by Hukseflux, measures soil heat flux, typically for energy-balance in flux systems. It is intended for applications requiring the highest possible degree of measurement accuracy. At least two sensors are required for each site to provide spatial averaging. Sites with heterogeneous media may require additional sensors.

Read More

Benefits and Features

  • Corrects for errors due to differences in thermal conductivity between the sensor and surrounding medium, temperature variations, and slight sensor instabilities
  • Compatible with most Campbell Scientific data loggers
  • Uses Van den Bos-Hoeksma self-calibration method to provide a high degree of measurement accuracy

Images

An example energy-balance installation where the HFP01SCs are installed with the CS616 and TCAV

Detailed Description

The HFP01SC consists of a thermopile and a film heater. The thermopile measures temperature gradients across the plate. During the in-situ field calibration, the film heater is used to generate a heat flux through the plate. The amount of power used to generate the calibration heat flux is measured by the datalogger. Each plate is individually calibrated, at the factory, to output flux.

Self-calibration corrects for errors due to differences in thermal conductivity between the sensor and surrounding medium, temperature variations, and slight sensor instabilities.

Specifications

Sensor Type Thermopile with film heater
Sensitivity 50 μV W-1 m-2 (nominal)
Nominal Resistance 2 Ω
Temperature Range -30° to +70°C
Expected Typical Accuracy ±3% of reading
Heater Resistance 100 Ω (nominal)
Heater Voltage Input 9 to 15 Vdc
Heater Voltage Output 0 to 2 Vdc
Duration of Calibration ±3 minutes @ 1.5 W (typically performed every 3 to 6 hours)
Average Power Consumption 0.02 to 0.04 W
Plate Diameter 80 mm (3.15 in.)
Plate Thickness 5 mm (0.20 in.)
Weight 200 g (7.05 oz) without cable

Compatibility

Please note: The following shows notable compatibility information. It is not a comprehensive list of all compatible products.

Dataloggers

Product Compatible Note
CR1000 (retired)
CR300
CR3000 (retired)
CR310
CR350
CR6
CR800 (retired)
CR850 (retired)

Downloads

CR1000X HFP01SC Example program v.2 (3 kb) 18-02-2020

CR1000X program that measures the HFP01SC, performs the self-calibration, and checks for calibration validity. Refer to the manual for the equations used for the self-calibration and calibration-validity checks. A table in the manual provides a cross reference of the terms in the equations in the manual with the constants and variables in the example data logger program.