The 108 is a rugged, accurate probe that measures temperature of air, soil, or water from -5° to +95°C. It easily interfaces with most Campbell Scientific dataloggers and can be used in a variety of applications.
Read MoreThe 108 consists of a thermistor encapsulated in an epoxy-filled aluminum housing. The housing protects the thermistor allowing the probe to be buried in soil or submerged in water.
Sensor Description | BetaTherm 100K6A1IA Thermistor |
Tolerance | ±0.2°C (over 0° to 70°C range) |
Temperature Measurement Range | -5° to +95°C |
Steinhart-Hart Equation Error | ≤ ±0.01°C over measurement range (CRBasic data loggers only) |
Polynomial Linearization Error | Typically < ±0.5°C over -5° to 90°C range (Edlog data loggers only) |
Time Constant in Air | 30 to 60 s (in a wind speed of 5 m s-1) |
Maximum Submergence | 15 m (50 ft) |
Maximum Cable Length | 305 m (1000 ft) |
Probe Diameter | 0.76 cm (0.3 in.) |
Probe Length | 10.4 cm (4.1 in.) |
Weight | 136 g (5 oz) with 3.05 m (10 ft) cable |
Please note: The following shows notable compatibility information. It is not a comprehensive list of all compatible products.
Product | Compatible | Note |
---|---|---|
CR1000 (retired) | ||
CR300 | ||
CR3000 (retired) | ||
CR310 | ||
CR350 | ||
CR6 | ||
CR800 (retired) | ||
CR850 (retired) |
One single-ended channel per probe is required; an excitation channel can be shared by several probes.
When exposed to sunlight, the 108 should be housed in a 41303-5A, 41303-5B, or RAD06 6-plate radiation shield. The louvered construction of these radiation shields allows air to pass freely through the shield, thereby keeping the sensor at or near ambient temperature. The shields’ white color reflects solar radiation.
The RAD06 uses a double-louvered design that offers improved sensor protection from driving rain, snow, insect intrusion and has lower self-heating in bright sunlight combined with higher temperatures (> 24°C [~75°F]) and low wind speeds (< 2 m/s [~4.5 mph]), giving a better measurement.
The 41303-5A and RAD06 attach to a crossarm, mast, or user-supplied pipe with a 2.5 to 5.3 cm (1.0 in to 2.1 in.) outer diameter. The 41303-5B attaches to a CM500-series pole or a user-supplied pole with a 5.1 cm (2.4 in.) outer diameter.
The 108 is suitable for shallow burial only. Placement of the sensor’s cable inside a rugged conduit may be advisable for long cable runs—especially in locations subject to digging, mowing, traffic, use of power tools, or lightning strikes.
The sensor can be submerged to 15 m (50 ft) or 21 psi. Please note that the 108 is not weighted. Therefore, the installer should either add a weighting system or secure the sensor to a fixed, submerged object, such as a piling.
To measure large numbers of probes, the AM16/32B multiplexer is recommended.