RETIRED ›
This product is no longer available and has been replaced by: CS547A-L. Some accessories, replacement parts, or services may still be available.
247-L Conductivity and Temperature Probe
Services Available
Repair No
Calibration No
Free Support No

Overview

The 247 was similar to the 247W-L, except it did not have a stainless steel sleeve, and therefore had a smaller diameter. The 247W measured conductivity using three passivated 316 stainless steel annular electrodes with dc isolation capacitors. A thermistor measured temperature.

Read More

Images

Detailed Description

The 247 had a pH range of 3.0 to 9.0, with a conductivity accuracy of ±5% (at 25°C) over 0.45 to 7.0 mS/cm range, ±10% (at 25°C) over 0.005 to 0.45 mS/cm range. Temperature polynomial linearization accuracy was typically less than ±0.1°C over 0° to +48°C range. Interchangeability accuracy was typically less than ±0.2°C over 0° to +50°C range.

The 247-L (pn 7343) was originally replaced with the CS547, which was replaced with the CS547A.

Specifications

  • Weight (w/10 ft cable): 0.2 kg (0.5 lb)
  • Diameter: 1.91 cm (0.75")
  • Length: 7.938 cm (3.125")
  • Operating Temperature: 0° to +50°C

Compatibility

The 247 required one differential analog input, one single-ended analog input, and two excitation channels.

Casos de estudio

Utah: Irrigation Network Monitoring
In an era when water is becoming a precious resource, the Emery Water Conservancy District......read more